Journal of Microwave Surgery
Online ISSN : 1882-210X
Print ISSN : 0917-7728
ISSN-L : 0917-7728
腎腫瘍
マイクロターゼ®を用いた腹腔鏡下腎部分切除術
加藤 司顯堀江 重郎奴田原 紀久雄東原 英二
著者情報
ジャーナル フリー

2003 年 21 巻 p. 123-127

詳細
抄録
Purpose : we performed laparoscopic partial nephrectomy in 8 cases. We will report the operative technique and the results.
Methods : After retroperitoneal cavity was expanded by balloon, which was replaced by a Hasson type trocar of 12 mm and two or three trocars were inserted. The tumor was separated and the coagulation line was set 5 to 8 mm apart from the tumor margin. The renal capsule was marked using an electric scalpel use at intervals of 5-8 mm. A needle electrode was inserted, and coagulation was done at 60-80 watt for 30-40 seconds and detachment for 10-15 seconds.
Results : The mean operation time was 170 minutes. The mean blood loss during operation was 71 mL. There was no major complication associated with laparoscopic procedure.
著者関連情報
© 2003 特定非営利活動法人 Microwave Surgery研究会
前の記事 次の記事
feedback