日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
原著
診断領域における散乱X線の可視化装置の製作と実習の提案
岸田 弥奈林 裕晃窪薮 友美竹上 和希井上 直花光 宏樹西原 貞光
著者情報
ジャーナル フリー

2013 年 69 巻 5 号 p. 500-507

詳細
抄録

We have built equipment that can visualize the angle distributions of scattered X-rays. The main body of the equipment is made of a lead-shielded box 340 mm long, 300 mm wide and 270 mm high. The collimated X-rays are introduced into the equipment from the front face, then scattered by the sample located in the center of the equipment. The X-rays scattering toward the upper side are detected by the phosphor plate. To verify the usability of the equipment, an experiment using diagnostic X-rays was carried out. X-rays with a tube voltage of 100 kV were narrowed down to 6 mmφ and a 2 mm-thick acrylic sample was irradiated. The experimental conditions with a tube current-time product of 300 mAs to 1500 mAs proved appropriate for obtaining suitable images on a 10 inch×12 inch phosphor plate. The obtained images were analyzed using ImageJ. The experimental values were in good agreement with the theoretical distribution calculated by Klein and Nishina. Because the distribution of the scattered X-rays can be visualized in relatively simple experiments using the developed equipment, it is hoped that it will be of use for the practical training of beginners.

著者関連情報
© 2013 公益社団法人 日本放射線技術学会