臨床技術
異なる線量計算法のバージョンアップに伴うカウチモデリングの最適化
桑原 潤一中田 学藤本 隆広中村 光宏佐々木 誠鶴田 裕輔矢野 慎輔東村 享治平岡 真寛
著者情報
ジャーナル フリー

2017 年 73 巻 1 号 p. 42-50

詳細
抄録

In external radiotherapy, the X-ray beam passes through the treatment couch, leading to the dose reduction by the attenuation of the couch. As a method to compensate for the reduction, radiation treatment planning systems (RTPS) support virtual couch function, namely “couch modeling method”. In the couch modeling method, the computed tomography (CT) numbers assigned to each structure should be optimized by comparing calculations to measurements for accurate dose calculation. Thus, re-optimization of CT numbers will be required when the dose calculation algorithm or their version changes. The purpose of this study is to evaluate the calculation accuracy of the couch modeling method in different calculation algorithms and their versions. The optimal CT numbers were determined by minimizing the difference between measured transmission factors and calculated ones. When CT numbers optimized by Anisotropic Analytical Algorithm (AAA) Ver. 8.6 were used, the maximum and the mean difference of transmission factor were 5.8% and 1.5%, respectively, for Acuros XB (AXB) Ver. 11.0. However, when CT numbers optimized by AXB Ver. 11.0 were used, they were 2.6% and 0.6%, respectively. The CT numbers for couch structures should be optimized when changing dose calculation algorithms and their versions. From the comparison of the measured transmission to calculation, it was found that the CT numbers had high accuracy.

著者関連情報
© 2017 公益社団法人 日本放射線技術学会