Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Quantification of Dopamine in Brain Microdialysates with High-Performance Liquid Chromatography–Tandem Mass Spectrometry
Dawei QIQian ZHANGWanhong ZHOUJingjing ZHAOBo ZHANGYunfei SHAZhiqing PANG
Author information
JOURNAL FREE ACCESS

2016 Volume 32 Issue 4 Pages 419-424

Details
Abstract

The dopamine level in the brain and the mesolimbick dopaminergic system are responsible for nicotine addiction. In the present study, extracellular dopamine in nucleus accumbens was collected by a brain microdialysis technique. Also a sensitive high-performance liquid chromatography–tandem mass spectrometry (HPLC-MS/MS) method with a gradient elution procedure was developed for a precise determination of dopamine in brain dialysate. The retention time of dopamine was about 11.32 min. The linear range was 20 – 1000 pg/mL. The limit of detection (LOD) and the limit of qualification (LOQ) were 5 and 20 pg/mL, respectively. The recovery ranged from 98.2 to 109.0%, and both the intra-RSD and inter-RSD were below <9%, respectively. The probe recovery for dopamine in this brain microdialysis experiment was about 25%. Finally, the dopamine concentrations in the rat brain microdialysates were determined, and the pharmacokinetics of extracellular dopamine in the brain nucleus accumbens after an intravenous injection of nicotine was successfully evaluated.