Use of Field-Portable X-Ray Fluorescence (FPXRF) Analyzer to Measure Airborne Lead Levels in Korean Workplaces
Nam-Soo KimJin-Ho KimKyu-Dong AhnByung-Kook Lee
Author information
JOURNAL FREE ACCESS

2007 Volume 49 Issue 6 Pages 493-498

Details
Abstract

We evaluated the possibility of applying field-portable x-ray fluorescence (FPXRF) analysis as a rapid, on-site and near real-time method for evaluating airborne lead contamination in Korean workplaces. A total of 287 airborne lead filter samples were measured in 12 lead-using workplaces during routine industrial hygienic monitoring procedures as required by Korean government regulations. All filter samples were collected using the standard industrial hygiene sampling protocol described in NIOSH Method 7300 using closed-face 37-mm cassettes with preloaded cellulose ester membrane filters with a pore size of 0.8 μm. The samples were first analyzed using the non-destructive, FPXRF analytical method (NIOSH method 7702), and then subsequently analyzed using inductively coupled plasma atomic emission spectrophotometry (ICP) (NIOSH method 7300) as a reference analytical method. Pair-wise comparison of filter samples using the paired t-test revealed no statistically significant differences between the two methods over a wide range of airborne lead levels (0.018-0.201 μg/m3) either over the industries assessed or separately in the 12 lead-using workplaces. Linear regression of the data between the ICP and FPXRF methods produced a slope of 1.03, a y-intercept of 0.13 μg/sample, and a coefficient of determinant (r2) of 0.975 for all the data. For samples in the range from 0 to 100 μg, the corresponding values were 1.07, -1.20 μg/sample, and 0.925, respectively. There were no significant differences in the regression analyses of the three industry types (r2=0.964-0.982). Our data suggest that FPXRF data are highly correlated with those from the laboratory-based ICP method in terms of accuracy, precision, and bias. Therefore, FPXRF can be used for the rapid, on-site analysis of lead air-filter samples for values up to 26 μg/sample prior to laboratory confirmation by the ICP method.

Content from these authors

This article cannot obtain the latest cited-by information.

2007 by the Japan Society for Occupational Health
Previous article Next article
feedback
Top