Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
Original Papers
Integration of Flow Injection Analysis and Zeptomole-Level Detection of the Fe(II)-o-Phenanthroline Complex
Kiyoshi SATOManabu TOKESHITakehiko KITAMORITsuguo SAWADA
Author information
JOURNALS FREE ACCESS

1999 Volume 15 Issue 7 Pages 641-645

Details
Abstract

Microchannels having a 150×100 µm cross section were fabricated in a quartz glass chip as a component in an integrated flow injection analysis (FIA) system. They were put to use for flow, mixing, reaction, and detection. The reaction system was a chelating reaction of divalent iron ion with o-phenanthroline (o-phen), and a photothermal microscope was applied for the ultra-sensitive detection of the non-fluorescent reaction product. Nano liter levels of solutions were introduced and transported by capillary action and mixed by molecular diffusion. Zeptomole levels of the reaction product were detected quantitatively. This was the first demonstration of an on-chip chemical determination device which integrates the primitive FIA system without using electroosmotic liquid control or fluorometric determination.

Information related to the author
© 1999 by The Japan Society for Analytical Chemistry