Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Construction and Performance Characteristics of New Fentanyl-Selective Plastic Membrane Electrode
Li Juan PENGMeng Liang WENYun YAO
Author information
JOURNAL FREE ACCESS

2001 Volume 17 Issue 7 Pages 815-818

Details
Abstract

The construction and performance characteristics of an ion-selective electrode for fentanyl-drug cation, based on an ion-pair complex with tetrakis[3,5-bis-(trifluoromethyl)phenyl]borate anion in a PVC matrix were studied. A linear response for 1 × 10-5 mol dm-3 to 1 × 10-2 mol dm-3 drug with a slope of 57.9 ± 0.5 mV/decade was established. The optimum pH range was 2 to 6. The lower detection limit was 6.29 × 10-6 mol dm-3 fentanyl citrate (2.1165 µg cm-3 fentanyl). There were negligible interferences from a number of inorganic cations, structural analogues, and some common drug additives in injections. The electrode proposed has been successfully applied to determine fentanyl citrate in injections. The results correlated well with those obtained by the United States Pharmacopoeia standard procedure.

Content from these authors
© 2001 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback