Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
In vitro and Short-term in vivo Characteristics of a Kel-F Thin Film Modified Glucose Sensor
Sun Kil KANGRan-A JEONGSejin PARKTaek Dong CHUNGSunmin PARKHee Chan KIM
Author information
JOURNAL FREE ACCESS

2003 Volume 19 Issue 11 Pages 1481-1486

Details
Abstract

A new outer layer composition, consisting of polytetrafluoroethylene (PTFE), Kel-F oil, and Nafion, is suggested to minimize the detrimental effect of dissolved oxygen and to extend the linear response range of a glucose oxidase(GOx)-based sensor using nonconducting polymer. The morphology of Kel-F/PTFE/Kel-F/Nafion polymeric laminate was followed during fabrication by SEM. When Kel-F film was formed on the PTFE outer layer, the linear response was extended to 21 mM, at a sensitivity of 2.8 ± 0.8 nA/mM mm2. We demonstrate that a sensor without Kel-F/PTFE/Kel-F/ Nafion outer layer is relatively oxygen dependent, whereas by comparison a sensor with Kel-F/PTFE/Kel-F/Nafion outer layer is oxygen independent. The current of such a glucose sensor implanted in the subcutaneous tissue stabilized within 60 min, and the lag between blood glucose changes and sensor output was within 1 min. The in vivo characteristics of the glucose sensor described show great promise for one-point in vivo calibration.

Content from these authors
© 2003 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top