2003 Volume 19 Issue 3 Pages 387-389
Two-color (1+1′) resonance-enhanced multiphoton ionization/time-of-flight mass spectrometry (REMPI/TOF-MS) combined with supersonic jet (SSJ) spectrometry has been demonstrated. The methodology uses a two-color laser beam consisting of a Stokes beam generated by stimulated Raman scattering (SRS) and a fundamental beam for the excitation and subsequent ionization of p-dichlorobenzene. The MPI signal was found to be substantially increased when a two-color laser beam was employed. This approach greatly simplifies the analytical system and reduces the cost of the instrumentation required for two-color ionization. The potential advantage of applying this method for the analysis of polychlorinated dibenzo-p-dioxin (PCDD) and polychlorinated dibenzofuran (PCDF) emitted from an incinerator is discussed.