A Miniaturized Fluorescence Detector with a Windowless Flow Cell Using a Light-Emitting Diode
Bingcheng YANGYafeng GUAN
著者情報
ジャーナル フリー

2003 年 19 巻 4 号 p. 633-634

詳細
抄録
A miniaturized fluorescence detector using a high-brightness light-emitting diode as an excitation source was constructed and evaluated. A windowless flow cell based on a commercial four-port cross fitting was designed to reduce the stray-light level and to eliminate the optical alignment. The observed detection limit for fluorescein was 26 nM in the continuous-flow mode. The error in the reproducibility of the responses was evaluated by the FIA method, and was found to be within 2% RSD.