Rapid Separation of Inorganic Anions by Capillary Ion Chromatography Using Monolithic Silica Columns Modified with Dilauryldimethylammonium Ion
Atsushi SUZUKILee Wah LIMToyohide TAKEUCHI
著者情報
ジャーナル フリー

2007 年 23 巻 9 号 p. 1081-1084

詳細
抄録
The rapid separation of inorganic anions was determined by capillary ion chromatography using monolithic silica capillary columns modified with dilauryldimethylammonium bromide. The stability of the modified stationary phase was satisfactory owing to a strong hydrophobic interaction between the lauryl groups of the reagent, even if the eluent did not contain dilauryldimethylammonium ion. Bromide in seawater samples could be determined by the present system. The repeatability of a retention time of bromide for six successive measurements was around 1.8% when a 500 mM sodium chloride aqueous solution was used as the eluent. Seawater samples were directly injected onto the prepared column without any interference of matrix ions, because an aqueous solution of high-concentration sodium chloride could be used as the eluent. Bromide in seawater samples could be determined within 2 min.
著者関連情報
© 2007 by The Japan Society for Analytical Chemistry