Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
An Optical Rotatory Detector for High-Performance Liquid Chromatography Using Polarization Modulation
Atsushi YAMAMOTOMio KAWAIMitsunori SAKAMOTOShuji KODAMAKazuichi HAYAKAWA
Author information
JOURNAL FREE ACCESS

2008 Volume 24 Issue 3 Pages 361-364

Details
Abstract

A sensitive and variable-wavelength optical rotatory (OR) detector for high-performance liquid chromatography is presented. This design is entirely different from that of conventional OR detectors consisting of a crossed polarizer pair. By placing a polarizing prism and a retardation plate into a commercial circular dichroism (CD) detector, the OR signal was obtained. The Mueller matrix approach was used to prove the principle of the OR signal appearance. Sugars and 4-androstene-3,17-dione were chosen as test compounds. The limit of detection was below 0.5 µg of injected sucrose at 260 nm, which was superior to that obtained with a conventional OR detector. For 4-androstene-3,17-dione, which is CD active, and shows a large anomalous OR dispersion curve, our detector gave a large OR signal with approximately half the intensity of the CD signal at 340 nm.

Content from these authors
© 2008 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top