Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Simultaneous Quantification of Iodine and Other Elements in Infant Formula by ICP-MS Following an Acid Digestion with Nitric Acid and Hydrogen Peroxide
Koji FUJISAKIHiroshi MATSUMOTOYukiko SHIMOKAWAKenji KIYOTAKI
Author information
JOURNAL FREE ACCESS

2016 Volume 32 Issue 2 Pages 167-170

Details
Abstract

A method for quantifying iodine in infant formula is described. Nitric acid and hydrogen peroxide converted iodine into iodate in microwave-assisted digestion and prevented iodine volatilization and memory effects. Acetic acid as a carbon source was added to both the sample and standard solutions as a countermeasure against carbon charge transfer to iodine and the addition of acetic acid helped to enhance the sensitivity. The instrument limit of quantification was 0.1 ng mL−1 and the relative standard deviation was less than 3%. The spike recoveries were between 94.8 and 106%. Good agreement with the values obtained using the tetramethylammonium hydroxide method was obtained for infant formula sold in several countries. This method permitted the simultaneous determination of iodine and 12 other important elements (Na, Mg, P, K, Ca, Cr, Mn, Fe, Cu, Zn, Se and Mo) in infant formula.

  Fullsize Image
Content from these authors
© 2016 by The Japan Society for Analytical Chemistry