On-line Analysis of Catalytic Reaction Products Using a High-Pressure Tandem Micro-reactor GC/MS
Atsushi WATANABEYoung-Min KIMAkihiko HOSAKAChuichi WATANABENorio TERAMAEHajime OHTANISeungdo KIMYoung-Kwon PARKKaige WANGRobert R. FREEMAN
Author information
JOURNAL FREE ACCESS

2017 Volume 33 Issue 9 Pages 1085-1089

Details
Abstract

When a GC/MS system is coupled with a pressurized reactor, the separation efficiency and the retention time are directly affected by the reactor pressure. To keep the GC column flow rate constant irrespective of the reaction pressure, a restrictor capillary tube and an open split interface are attached between the GC injection port and the head of a GC separation column. The capability of the attached modules is demonstrated for the on-line GC/MS analysis of catalytic reaction products of a bio-oil model sample (guaiacol), produced under a pressure of 1 to 3 MPa.