Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Evaluation of Relative Sensitivity Factors for Elemental Analysis of Aluminum and Magnesium Using Glow Discharge Mass Spectrometry with a Fast-Flow Grimm-type Ion Source
Kentaro YAMADAMichihisa UEMOTO
Author information
JOURNAL FREE ACCESS

2018 Volume 34 Issue 6 Pages 743-746

Details
Abstract

Relative sensitivity factors on glow discharge mass spectrometry were evaluated for unalloyed and alloyed metals of aluminum and magnesium using a glow discharge mass spectrometer operated with fast flow Grimm-type source. All the elements measured could be classified into two groups, i.e. a group of elements could be determined with repeatability equal to or less than 15% relative standard deviation, while another group could be determined with RSDs of greater than 15%. The latter is mainly due to the instability of discharge condition and elemental segregation onto certified reference materials.

Fullsize Image
Content from these authors
© 2018 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top