Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Room-temperature Chelate Vapor Generation of Lead Using Ammonium O,O-Diethyl Dithiophosphate as a Chelating Reagent and Determination by Atomic Fluorescence Spectrometry in Environmental Water Samples
Shuai LIUJinsheng SUNJianzhong HUOXuchuan DUANYue LIU
著者情報
ジャーナル フリー
電子付録

2019 年 35 巻 5 号 p. 499-504

詳細
抄録

A novel gas-liquid separator (GLS) system for chelate vapor generation (Che-VG) combined with AFS was developed for the determination of trace Pb. It was shown that Pb can form a volatile chelate by mixing of Pb with ammonium O,O-diethyldithiophosphate (DEDTP) in various aqueous solutions. Under the optimal conditions (frit pores of the GLS, 5 – 15 μm; solution pH, 6.7; DEDTP concentration, 0.4%; flow rate of the two feed solutions in the on-line mode, 1.2 mL min−1; and carrier gas flow rate, 150 mL min−1), the calibration curve was linear up to 100 ng mL−1 Pb. The limit of detection (LOD) was 1.1 ng mL−1. The relative standard deviation was 5.6% for eleven replicate determinations of 10 ng mL−1 Pb. The efficiency of Che-VG was estimated to be 12%, and the volatile Pb species was preliminarily studied by ICP-OES. This method was applied to determine trace Pb in water samples.

Fullsize Image
著者関連情報
© 2019 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top