3D-Printed Microfluidic Nanoelectrospray Ionization Source Based on Hydrodynamic Focusing
Yu ZHAOShichang JIANGYuna BAIXueying HUANGBo XIONG
Author information
JOURNAL FREE ACCESS
Supplementary material

2021 Volume 37 Issue 6 Pages 897-903

Details
Abstract

Nanoelectrospray ionization (nESI) mass spectrometry (MS) is an ideal detection method for microfluidic chips, and its performances depend on nESI emitters. However, the fabrication of monolithic nESI emitters in chips was difficult. Herein, we propose a three-dimensional (3D) printing method to develop a microfluidic nanoelectrospray ionization source (NIS), composed of a nESI emitter and other components. Firstly, the NIS was compatible with a 50 – 500 nL min−1 nanoflows by imposing 3D hydrodynamic focusing to compensate for the total flow rate, achieving a 7.2% best relative standard deviation in the total ion current (TIC) profiles. Additionally, it was applied to probe thirteen organic chemicals, insulin, and lysozyme with adequate signal-to-noise ratios and an accuracy of m/z between 9.02 × 10−1 and 1.48 × 103 ppm. Finally, the NIS achieved comparable limits of detection compared with its commercial counterpart. Considering the standardized preparation of NIS, it would be a potential option to develop 3D-printed customized Chip-MS platforms.

Fullsize Image
Content from these authors
© 2021 by The Japan Society for Analytical Chemistry