Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Determination of Nickel, Cobalt, Copper, Thorium and Uranium in High-Purity Zinc Metal by ICP-MS with On-Line Matrix Separation
Masamitsu FUKUDAYutaka HAYASHIBEYasumasa SAYAMA
Author information
JOURNAL FREE ACCESS

1995 Volume 11 Issue 1 Pages 13-16

Details
Abstract

Traces of Co, Cu, Ni, Th and U in high-purity zinc metal were determined by inductively coupled plasma mass spectrometry (ICP-MS) in combination with flow-injection (FI) on-line matrix separation (FI-ICP-MS). The anion-exchange separation method of an HCl system was applied to the separation of Ni, Co, Cu, Th and U from zinc matrix. The matrix element, zinc, was adsorbed on an anion-exchange (BIO•AD AGl-X8) mini-column (2.0mm i.d.×300mm bed length), while the analytes were directly introduced into the ICP-MS. The following detection limits (in ng g-1) were obtained: Ni, 3.1; Co, 1.2; Cu, 4.0; Th, 0.12 and U, 0.48. The reproducibility has proved to be satisfactory with a relative standard deviation of less than 5% (at the 10ng ml-1 level, n=3). The method was successfully applied to determining trace impurities in three 6 and 7 nines grade of high-purity zinc metals and in three standard reference materials of high-purity zinc metal samples (from NIST).

Content from these authors
© The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top