Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Electrochemical Detector Using Carbon Fiber Working Electrode for Micro High Performance Liquid Chromatography
Masashi GOTOKen SHIMADAToyohide TAKEUCHIDaido ISHII
Author information
JOURNAL FREE ACCESS

1988 Volume 4 Issue 1 Pages 17-20

Details
Abstract

A tubular electrolytic cell of ultra micro volume, made of a single carbon fiber (7μm o.d., 15mm length) as the working electrode material and a fused silica tube (50μm i.d.) as the flow cell part, was designed for use as the electrochemical detector in micro high performance liquid chromatography (micro HPLC). The effective cell volume is about 0.03μl. This detector with carbon fiber was useful for detection of catecholamines and their metabolites. Their detection limits were about 1-2pg and the linear dynamic range was about 103. The detector was successfully applied to urinary catecholamine analysis by micro HPLC and to rapid detection of electroactive species in fast micro HPLC.

Content from these authors
© The Japan Society for Analytical Chemistry
Previous article Next article
feedback