Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
CONTINUOUS AUTOMATED DETERMINATION OF AMINO ACIDS BY ENZYME ELECTRODE POTENTIOMETRY
HEUNG LARK LEESEUNG-TAE YANGHO-JIN SOHN
Author information
JOURNAL FREE ACCESS

1991 Volume 7 Issue Supple Pages 293-295

Details
Abstract

Continuous automated system with enzyme reactor, gas dialyzer, and tubular PVC-nonactin membrane ammonium ion-selective electrode has been developed for the determination of amino acids. The enzyme reactor was constructed to the packed-bed form which filled with nylon-6 particle in the teflon tube. The linear concentration ranges of determination of L-asparagine and L-glutamine in physiological water were 1.0×10-5-2.8×10-3M and 6.5×10-5-2.4×10-3M, respectively.