Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Comparison of Avidin and Ovomucoid as Chiral Selectors for the Resolution of Drug Enantiomers by High-Performance Liquid Chromatography
Yoshiya ODANariyasu MANONaoki ASAKAWAYutaka YOSHIDATadashi SATOTerumichi NAKAGAWA
Author information
JOURNAL FREE ACCESS

1993 Volume 9 Issue 2 Pages 221-228

Details
Abstract

The effects of the pH, organic modifiers and buffer salt of the mobile phase upon the retention and separation characteristics of avidin- and ovomucoid-bonded silica columns as proteinous chiral stationary phases for HPLC were investigated. The enantioselectivity was greatly affected by changing the mobile-phase composition, possibly because of induced alterations in the protein conformation. Hydrophobic and Coulombic interactions play major roles in the retention, and there are some similarities between the avidin column and the ovomucoid column. However, the stereoselectivity does not necessarily reflect the retention processes on these columns. No significant decrease in the retention or enantioselectivity was observed, even after the injection of 200 samples.

Content from these authors
© The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top