EVALUATION OF ENZYME MULTIPLIED IMMUNOASSAY TECHNIQUE GENTAMICIN ASSAY
COMPARISON WITH BIOASSAY
HIROSHI SAKATAMASATOSHI TAKIMOTOYASUYUKI FUJITATAKESHI KITSUTAKA
Author information
JOURNAL FREE ACCESS

1982 Volume 35 Issue 5 Pages 1229-1232

Details
Abstract
To assess an enzyme multiplied immunoassay technique (EMIT) for determination of gentamicin, comparison was done with a microbiological bioassay. Twenty-nine samples from 5 patients and 1 volunteer were assayed by both EMIT and bioassay methods.
Concentrations of gentamicin obtained with EMIT were compared with those with bioassay.Statistical calculation resulted in a regression linewith a correlation coefficient of 0.9496, a slope of 0.735 and an ordinate intercept of 0.878.
To test the specificity of EMIT, serum controls which contain a constant quantity of gentamicin and a variable quantity of ampicillin concomitantlywere assayed for gentamicin concentrations by both EMIT and bioassay.This test demonstrated that EMIT is free from interference of ampicillin while bioassay is vulnerable to interference.
To conclude, EMIT proved to be a practical alternative to current bioassays for determination of serum gentamicin concentrations.