AUDIOLOGY JAPAN
Online ISSN : 1883-7301
Print ISSN : 0303-8106
ISSN-L : 0303-8106
Comparison Study of Hearing Levels Assessed by Automated DPOAE and Pure Tone Audiometry
Yasuko ShibataHarumi AraoHiroyuki NakayamaTosie KatouTakao MorikawaKatumi Asami
Author information
Keywords: Automated DPOAE
JOURNAL FREE ACCESS

2004 Volume 47 Issue 1 Pages 56-62

Details
Abstract

The aim of this study is to evaluate an Automated DPOAE and to compare the result of Automated DPOAE with pure tone hearing levels.
122 ears of 61 children between 19 and 179 months of age were examined by utilizing the Automated DPOAE (ERO SCAN).
33 ears passed the screening. 97% of the passed ears had audiometric thresholds within 40dB for 2kHz and 100% for 4kHz. On the other hand, one ear which showed moderate sensorineural hearing loss with 40dB hearing level for 4kHz passed the screening.
Automated DPOAE is useful for hearing screening, especially for the high frequency range. In addition, we have to pay much composite attention also to language development and articulation, and ongoing audiometry should be carried out if there is any doubt concerning hearing loss.