Bulletin of the Chemical Society of Japan
Online ISSN : 1348-0634
Print ISSN : 0009-2673
ISSN-L : 0009-2673
The Determination of Tin in Tin Metal and in Zircaloy by Precise Coulometric Titration
Takayoshi YoshimoriIkuko MatsubaraTatsuhiko TanakaKazuo YoshidaKazuo TanakaToyokazu Tanabe
Author information
JOURNALS FREE ACCESS

1971 Volume 44 Issue 3 Pages 734-737

Details
Abstract

The purities of tin metals are determined directly by precise coulometric titration. The sample is dissolved with hydrochloric acid in a nitrogen atmosphere. The solution is then introduced into a hot antimony column. The reduced tin(II) ions are titrated with electrolytically-generated iodine. The titration is accomplished in a special cell by the use of which any disturbance of atmospheric oxygen is avoided. The titration efficiency under these acidic media was decreased when the current density was increased more than several mA/cm2. The results obtained by two different analysts for the sample of the pure tin metal (5-nine) were 99.996 and 99.991%. The standard deviations of the results were 0.025 and 0.016% respectively. The tin in the other tin metals and in Zircaloy samples was determined satisfactorily without the use of any experimental factor.

Information related to the author

This article cannot obtain the latest cited-by information.

© The Chemical Society of Japan
Previous article Next article
feedback