Detection of tetramethylthiuram disulfide and related compounds using the chemiluminescence of tris(2,2'-bipyridyl)ruthenium(III) cation
Nobumitsu NIINAHitoshi KODAMATANIKeiitsu SAITOShigeo YAMAZAKI
Author information
JOURNAL FREE ACCESS

2003 Volume 52 Issue 9 Pages 763-767

Details
Abstract

The chemiluminescence (CL) of tris(2,2'-bipyridyl)ruthenium(III) cation was applied to the determination of tetramethylthiuram disulfide (TMTD). In this work, a sample was injected to an HPLC system with Chromolith Performance RP-18e. The elueted sample and CL reagent were mixed, and sent to a CL-detector. The detection limit of TMTD using the HPLC system was 0.15 pmol (S/N = 3). The calibration curve was linear over the injection amount range from 0.6 to 20 pmol. To improve the sensitivity of the CL method, the water that generated the background was removed from the system. As a result, the sensitivity was improved.

Content from these authors
© The Japan Society for Analytical Chemistry 2003