Electrochemistry
Online ISSN : 2186-2451
Print ISSN : 1344-3542
ISSN-L : 1344-3542
Article
Potentiometric Titration Based on the Reference Electrode Equipped with Ionic Liquid Salt Bridge — 1. Precipitation Titration of Chloride with Silver Ions in Water
Takashi KAKIUCHI Ryunosuke TANIGOAtsushi TANITakeshi YAMAZAKIKohta KOMATSUBARAKeiji NAKANOMasahiro YAMAMOTO
著者情報
ジャーナル オープンアクセス HTML
J-STAGE Data

2023 年 91 巻 1 号 p. 013001

詳細
抄録

A reference electrode equipped with ionic liquid salt bridge consisting of tributyl(2-methoxyethyl)phosphonium bis(pentafluoroethanesulfonyl)amide has been employed for potentiometric precipitation titration of chloride with silver ions in water at 25 °C. A model for the titration curve was regressed to experimental curves, taking into account the change in the activity coefficients of relevant ionic species in the course of the titration, to obtain the least square estimates of two adjustable parameters in the model, the solubility product (Ksp) and the analyte concentration. The least-square estimate of Ksp, (1.840 ± 0.060) × 10−10, i.e., pKsp = 9.736 ± 0.014, is in good agreement with literature data, but with higher precision.

著者関連情報
© The Author(s) 2022. Published by ECSJ.

This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 License (CC BY, http://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse of the work in any medium provided the original work is properly cited. [DOI: 10.5796/electrochemistry.22-66119].