Electrochemistry
Online ISSN : 2186-2451
Print ISSN : 1344-3542
ISSN-L : 1344-3542
Regular Papers
Wearable Microfluidic Dual Screen-Printed Sensor for Simultaneous Monitoring of Sweat pH and Lactate
Isao SHITANDA Momo CHIBANoya LOEW Hidefumi YOSHIZAWAMasahiro MOTOSUKETakahiro MUKAIMOTOShinya YANAGITATatsunori SUZUKITsutomu MIKAWAHikari WATANABEMasayuki ITAGAKI
著者情報
ジャーナル オープンアクセス HTML

2025 年 93 巻 11 号 p. 117001

詳細
抄録

Sweat lactate monitoring has recently garnered increasing attention for assessing athlete performance, optimizing training, and determining fatigue levels. Thus, this study introduces a wireless dual-biosensor device for continuous monitoring of sweat lactate and pH. The pH sensor exhibits a sensitivity of −61 ± 0.5 mV pH−1, while the lactate sensor displays a sensitivity of 0.0149 ± 0.0004 mA cm−2 mM−1 at pH 7.5. We establish a method to precisely measure lactate within the pH range of 5.5 to 7.5, taking into account the pH-dependent sensor sensitivity. The dual-biosensor device incorporates polydimethylsiloxane microfluidics for collection of sweat, its delivery to the sensor, and disposal of used sweat, enabling continuous and precise monitoring. On-body tests demonstrate the ability of the proposed device to continuously monitor sweat pH and lactate levels in individuals during exercise.

Fullsize Image
著者関連情報
© The Author(s) 2025. Published by ECSJ.

This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 License (CC BY, https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse of the work in any medium provided the original work is properly cited. [DOI: 10.5796/electrochemistry.25-00108].
https://creativecommons.org/licenses/by/4.0/
次の記事
feedback