IEICE Electronics Express
Online ISSN : 1349-2543
ISSN-L : 1349-2543
RF power measurement in D-band using down-converter calibrated by three-mixer method
Katsumi FujiiToshihide TosakaKaori FukunagaYasushi Matsumoto
Author information
JOURNAL FREE ACCESS

2012 Volume 9 Issue 13 Pages 1096-1101

Details
Abstract

To establish an RF power standard above 110GHz, we have developed a measurement system using a down-conversion mixer with a spectrum analyzer, and the conversion loss of the mixer was determined by the three-mixer method. This method measures the three transmission S-parameters (S21) of three mixer pairs using a vector network analyzer. The experimental results proved that the three-mixer method can determine the conversion loss of mixers with an uncertainty of 0.5dB (level of confidence: 95%) in the entire range of the D-band. Using this uncertainty and the uncertainty of the spectrum analyzer, the proposed system can measure the RF power in the D-band with an uncertainty of less than 2.0dB.

Content from these authors
© 2012 by The Institute of Electronics, Information and Communication Engineers
Previous article Next article