岩石鉱物科学
Online ISSN : 1349-7979
Print ISSN : 1345-630X
ISSN-L : 1345-630X
原著論文
誘導結合プラズマ質量分析法による岩石試料の微量元素の定量分析
矢嶋 一仁藤巻 宏和
著者情報
ジャーナル フリー

2002 年 31 巻 5 号 p. 249-260

詳細
抄録
For quantitative analyses of trace elements in rock samples by inductively coupled plasma-mass spectrometry (ICP-MS), it is necessary to accommodate instrumental drift and irregularities in detector linearity. Interferences of oxide molecular ions and non-spectral matrix effects are also taken into consideration to establish an appropriate analytical method. Relatively rare trace elements in rock samples, such as rare earth elements, are determined by alternatively analyzing a 10-ppb standard solution and 1000 times diluted rock solutions containing a 10-ppb Indium internal standard. Analytical results for both samples and standards are normalized to the internal standard, which enables minimization of instrumental drift and correction of uncertainty in dilution factors. Using the internal standard, matrix effect is effectively canceled out at the dilutions factor. Determinations of elements present in relatively high concentrations, such as Rb, Sr, Y and Zr, is achieved by increasing dilution to achieve acceptable detector linearity. This further suppresses matrix effects originating from concomitant major elements.
著者関連情報
© 2002 日本鉱物科学会
次の記事
feedback
Top