Special Issue Paper
Electromagnetic Compatibility of Transcutaneous Energy Transmission System
for Totally Implantable Artificial Heart
Kenji ShibaKohji Koshiji
Author information
JOURNAL FREE ACCESS

2003 Volume 123 Issue 7 Pages 1219-1227

Details
Abstract

Transcutaneous Energy Transmission (TET) is one way of providing the energy needed to power a totally implantable artificial heart (TIAH). In the present study, an externally coupled TET system was implanted in a prototype human phantom to evaluate emission and immunity. In the emission evaluation, measurements were conducted based on CISPR Pub.11 and VDE 0871 standards, while immunity tests were based on the standards of the IEC 61000-4 series.
The magnetic field of the radiated emission was measured using a loop antenna. At 0.1[MHz], we found the greatest magnetic field of 47.8 [dBμA/m], somewhat less than CISPR’s upper limit of 54 [dBμA/m]. For the conducted emission, by installing a noise filter and ferrite beads in the input section of the DC-power supply, conducted emission could be kept within the allowable limits of CISPR Pub.11 and VDE 0871. Finally, the immunity tests against radiated and conducted emission, electrostatic discharge and voltage fluctuation proved that the prototype could withstand the maximum level of disturbance.
These results confirmed that the TET system implanted in a human phantom could, through modification, meet the emission and immunity standards.