電気学会論文誌C(電子・情報・システム部門誌)
Online ISSN : 1348-8155
Print ISSN : 0385-4221
ISSN-L : 0385-4221
<神経工学 (計測技術)>
低導電性保護膜を有する光アドレス電極の光電気特性と同電極を用いた電気刺激の空間分解能
硯川 潤高橋 宏知神崎 亮平中尾 政之神保 泰彦
著者情報
ジャーナル フリー

2007 年 127 巻 10 号 p. 1581-1587

詳細
抄録

Conventional microelectrode arrays (MEAs) cannot always access desired neurons due to low electrode density and small electrode number. To overcome this problem, we have proposed and developed a light-addressable planar electrode on a glass substrate. The electrode has a 3-layer structure, namely a transparent SnO2 layer, an hydrogenated amorphous silicon (a-Si:H) layer, and a low-conductive passivation layer. Illumination to the a-Si:H layer increases the conductivity of a-Si:H and generates a virtual electrode at the surface of the illuminated site. In the present study, we evaluated the photoelectric property of the developed electrode and estimated the spatial resolution of the light-addressed stimulation. Illumination to the electrode increased stimulus intensity by up to 60-folds. This illumination-induced intensity change sufficiently followed high-frequency illumination switching. The simultaneous fluo-4 Ca2+-imaging, thus, successfully monitored post-stimulus fluorescence transients by instantaneously shutting out the excitation light during stimulus pulse application. By monitoring stimulus induced responses of cell aggregations, we estimated the spatial resolution of the light-addressed stimulation at 10 μm or more with an addressing illumination spot of 70 μm in diameter.

著者関連情報
© 電気学会 2007
前の記事 次の記事
feedback
Top