アナログ順フィルタとディジタル逆フィルタを用いた相補的な帯域拡張法と非接触心電計測への応用
小和田 大輔植野 彰規
著者情報
ジャーナル フリー

2011 年 131 巻 11 号 p. 1966-1974

詳細
抄録

In this paper, we propose a novel method for extending bandwidth of ECG measuring device without deteriorating the baseline tolerance for body motion. The proposed method is realized by synthesizing a two-stage analog forward filter with two different corner frequencies and a digital inverse filter having a corner frequency identical with the higher one of the analog filter. We applied the method to bed-type capacitive ECG sensor which can detect electrocardiographic potential non-intrusively and indirectly. The results demonstrated that the proposed method could precisely recover low-frequency components like T-wave. Furthermore, we confirmed that QTc (corrected QT interval) could be estimated from the recovered wave and that the QTc correlated 0.81 on the average with that obtained from Lead II ECG for seven subjects. These results indicate that the proposed method is useful to screening of long QT syndrome by combining the method with the bed-type capacitive ECG sensor.

著者関連情報
© 電気学会 2011
前の記事 次の記事