2011 年 131 巻 9 号 p. 1577-1586
This study introduces a novel technique for estimating visual field by using overlapped fixation patterns obtained by amounts of displacement in voluntary eye movements during searching-tracking trials, as an alternative screening medical device for visual field examination. Proposed system was evaluated by glaucoma patients, in order to study whether or not the proposed system can be used as a visual field detection device for screening. As a result, the proposed system detected the visual field abnormality to a certain extent. However, there were some cases that detection of Mariotte blind spots was inaccurately performed. The experimental results revealed that there was room to reconsider our understanding regarding the effect of ptosis, overlapped with eye sight as well as calibration of the display position to the correct location to be examined. Results of the evaluation experiment indicated that this proposed system has a potential to be used as a visual field examination device for screening.