電気学会論文誌C(電子・情報・システム部門誌)
Online ISSN : 1348-8155
Print ISSN : 0385-4221
ISSN-L : 0385-4221
<生体医工学・福祉工学>
MRIのためのエレベータに起因する変動磁場を低減するシャフト壁の遮蔽方法の検討
鎌田 清孝内門 大地玉利 陽三湯ノ口 万友
著者情報
ジャーナル フリー

2017 年 137 巻 2 号 p. 342-347

詳細
抄録

This paper describes an optimal shielding shape of elevator shafts for installations of MRI and Electron Beam Lithography System. The imaging of the MRI is mainly disturbed by the magnetic flux density fluctuations (MFDF) in the direction that the uniform flux is applied with amplitudes of 1.5 T and 3.0 T. Almost all MRIs are set horizontally, so the shielding method for the horizontal component of the MFDF was primarily investigated. A shielding shaft divided into upper and lower parts with a gap also enables the opening for going in and out from the elevator. Firstly, the MFDFs due to the movement of the elevator were measured. Then, the magnitudes and directions of the magnetic moments that can model the magnetic materials of the elevator were estimated using an inverse problem, so that the MFDFs obtained agree with those measured. Secondly, the relationship between shape of shielding shaft and the estimated fluctuations of leakage magnetic field in the horizontal and vertical directions due to the movement of the modeled elevator was investigated using a 3D finite linear magnetic-field analysis. Lastly, the optimum shape of the elevator-shielding shaft for preventing MFDFs due to the movement of the elevator was investigated by analysis.

著者関連情報
© 2017 電気学会
前の記事 次の記事
feedback
Top