時間分解分光法によるヘモグロビン濃度の推定とBOLD信号との定量的比較
庭野 恭彰赤尾 旭彦四本 裕子神保 泰彦小谷 潔
著者情報
ジャーナル フリー

2018 年 138 巻 7 号 p. 790-798

詳細
抄録

Time-Resolved Near-Infrared Spectroscopy (TR-NIRS) is a method of analyzing the time response of the reflected light using a light diffusion equation, and enables us to measure the hemoglobin (Hb) absolute concentration, In this research, we evaluated the applicability of TR-NIRS for the measurement of brain activity. First, we conducted the simultaneous measurement of Continuous Wave Near-Infrared spectroscopy (CW-NIRS) and TR-NIRS in the working memory tasks. The results show the conventional analysis is not applicable, because the fluctuation is larger than the task-related change. Second, we proposed a new signal processing method for suppressing the fluctuation and examined the validity in the original fantom. Finally, by using the proposed method, we conducted the simultaneous measurement of TR-NIRS and functional MRI (fMRI) in the working memory tasks, and verify depth of reach and light path of near infrared light. The result suggests that we can estimate the Hb concentration due to brain activity using the proposed method.