Resovist®および感温磁性体の混合物の磁気特性を利用したワイヤレス温度計測・誘導加熱システムの研究
安藝 史崇Tonthat Loi齊藤 元吉村 昇水戸部 一孝
著者情報
ジャーナル 認証あり

2019 年 139 巻 1 号 p. 38-44

詳細
抄録

The aim of this paper is to clarify the treatable distance of Ferromagnetic Implant with Low Curie Temperature (FILCT), superparamagnetic nanoparticles Resovist® (Resovist) and the mixture of FILCT and Resovist (Reso-FILCT) as heating elements by using a wireless temperature measurement induction heating system for hyperthermia under a constant temperature environment of 37℃. In this experiment, a high-frequency magnetic field was applied to the three heating elements and their temperature and the superimposed voltage were measured every 1.0 cm up to 6.0 cm until the temperature of heating element no longer exceeds 45℃ even after 300 seconds passed. As a result, it was clarified that the treatable distance of FILCT is within 1.0 cm, the temperature of Resovist is difficult to control, and the treatable distance of Reso-FILCT is within 5.0 cm. Furthermore, it was revealed that Reso-FILCT has an initial temperature rising rate of 3.6 times and a change in superimposed voltage of 1.28 times than that of FILCT. From this point, we considered that wireless temperature measurement induction heating system using Reso-FILCT as heating element is useful for clinical application.