医療用CMOS撮像モジュールへの応用を目指した撮像素子WL-CSPプロセスの開発
藤森 紀幸五十嵐 考俊下畑 隆博巣山 拓郎吉田 和洋中川 悠輔中村 力佐藤 敏郎
著者情報
ジャーナル フリー

2017 年 137 巻 2 号 p. 48-58

詳細
抄録

A new packaging technique has been realized and applied to the imaging module for medical endoscope. In order to realize minimal invasiveness for in-vivo medical device applications, the WL-CSP (Wafer Level-Chip Size Package) is one of the most attractive technology. However, in the field of medical device packaging, the packaging needs are diversified and required number of each package does not meet the huge sized wafer processing. In this new technique, the CMOS image sensor wafer is diced at first, and then individual image sensor chips are rearranged on another smaller handling wafer. As a result, suitable size of the handling wafer can be selected for the production volume, and lower production cost and shorter production lead time can be achieved. The packaged image sensor with new packaging technique showed no significant degration of charactristic of image sensor.

著者関連情報
© 2017 電気学会