International Heart Journal
Online ISSN : 1349-3299
Print ISSN : 1349-2365
ISSN-L : 1349-2365
Clinical Studies
Revival of the Forgotten
Temperature-Controlled Catheter Ablation of the Pulmonary Veins Using the Novel DiamondTemp Ablation System
Laura RottnerFabian MoserJulia MoserRuben SchlebergerMarc LemoinePaula MünklerLeon DinshawPaulus KirchhofFeifan OuyangAndreas RilligAndreas MetznerBruno Reissmann
Author information
JOURNAL FREE ACCESS

2022 Volume 63 Issue 3 Pages 504-509

Details
Abstract

Pulmonary vein (PV) isolation (PVI) by continuous, transmural and durable lesions is decisive for ensuring long-term freedom from atrial fibrillation (AF). AF ablation requires irrigated tip catheters to reduce thromboembolic complications. This precluded temperature-controlled delivery of radiofrequency (RF) energy.

The aim of this study was to evaluate feasibility, acute efficacy, and safety of an irrigated, temperature-controlled ablation catheter [DiamondTemp™ (DT) Medtronic®] for PVI.

Consecutive patients with AF underwent PVI using the DT catheter combined with high-power short-duration RF applications. Ablation settings were (1) a catheter tip temperature limit of 60°C, (2) a temperature-controlled power of 50 W, and (3) application duration of 10 seconds. The primary endpoint was acute isolation of PVs, reassessed after a 30-minute waiting period. Secondary endpoints included procedural parameters (defined as a catheter tip temperature of 50°C > 3 seconds, an impedance drop of 5-10 Ω) and the occurrence of serious adverse events.

Fifty consecutive patients [mean age 66 ± 12 years, 38 (76%) women, 24 patients with paroxysmal AF (48%)] were included. Median procedure and left atrial dwell time was 89 [68; 107] and 63 [52; 79] minutes, respectively. Mean number of RF applications was 59 ± 20, and mean total RF duration was 14 ± 6 minutes. Acute PVI was achieved in all patients solely using DT ablation. Acute PV reconnection within the waiting period occurred in five patients; all reconnected PVs were successfully reisolated. One major complication occurred.

In this study, the DT ablation system demonstrated high acute efficacy for PVI. Temperature-controlled ablation in conjunction with high-power short-duration applications might be effectively supported.

Content from these authors
© 2022 by the International Heart Journal Association