System Identification Method for Non-Invasive Ultrasound Theragnostic System Incorporating Mechanical Oscillation Part
Norihiro KoizumiKouhei OotaDongjun LeeHiroyuki TsukiharaAkira NomiyaKiyoshi YoshinakaTakashi AzumaNaohiko SugitaYukio HommaYoichiro MatsumotoMamoru Mitsuishi
著者情報
ジャーナル オープンアクセス

2014 年 8 巻 1 号 p. 110-119

詳細
抄録

In this paper, we propose a method for identifying systems incorporating a mechanical oscillation part for a non-invasive ultrasound theragnostic system (NIUTS). The NIUTS tracks and follows movement in an area requiring treatment (renal stones, in this study) by irradiating the area with high intensity focused ultrasound (HIFU). Blur noise caused by oscillation of the mechanical system adversely affects the servo performance. To solve this problem and enhance the servo performance, it is first necessary to identify those parts of the NIUTS system that incorporate a mechanical oscillation part. Secondly, we implemented a mechanical oscillation suppression filter based on the above-mentioned method for identifying the mechanical oscillation part.