医科器械学
Online ISSN : 1881-4875
Print ISSN : 0385-440X
マイクロコンピュータとの接続を可能とした尿流計
秦野 直浅野 友彦中島 史雄山本 秀伸篠田 正幸萩原 正通
著者情報
ジャーナル フリー

1985 年 55 巻 3 号 p. 107-110

詳細
抄録

In the field of urodynamics, microcomputers have become used for various data processing.This report introduced a uroflowmeter for on-line data processing. In this case, a higher precision uroflowmeter was required and this type uroflowmeter has been developed. In this flowmeter, sample and hold circuits were used for the weight-to-flow converter. As a result, a low drift and high stability of the flow curve were obtained. Further the measurement of the flow rate was able to started within 6 seconds after the power was turned on. The urine volume was calculated by a microcomputer from the flow curve obtained, and error rates in this case remained less than 2 % mostly.

著者関連情報
© 1985 一般社団法人日本医療機器学会