Rate-dependent Loss of Capture during Ventricular Pacing
Jingfeng WangHaiyan ChenYangang SuJunbo Ge
Author information
JOURNAL OPEN ACCESS

2015 Volume 54 Issue 19 Pages 2449-2451

Details
Abstract

A 63-year-old patient who had undergone atrial septal defect surgical repair received implantation of a single chamber VVI pacemaker for long RR intervals during atrial fibrillation. One week later, an intermittent loss of capture and sensing failure was detected at a pacing rate of 70 beats/min. However, a successful capture was observed during rapid pacing. Consequently, the pacing rate was temporarily adjusted to 90 beats/min. At the 3-month follow-up, the pacemaker was shown to be functioning properly independent of the pacing rate. An echocardiogram showed that the increased pacing rates were accompanied by a reduction in the right ventricular outflow tract dimension. The pacemaker was then permanently programmed at a lower rate of 60 beats/min.