Internal Medicine
Online ISSN : 1349-7235
Print ISSN : 0918-2918
ISSN-L : 0918-2918
CASE REPORTS
False Elevation of the Blood Tacrolimus Concentration, as Assessed by an Affinity Column-mediated Immunoassay (ACMIA), Led to Acute T Cell-mediated Rejection after Kidney Transplantation
Momoko KonoJumpei HasegawaHina OgawaKanae YoshikawaAyumi IshiwatariSachiko WakaiKazunari TanabeHiroki Shirakawa
Author information
JOURNAL OPEN ACCESS

2018 Volume 57 Issue 9 Pages 1265-1268

Details
Abstract

Tacrolimus is the most commonly used immunosuppressant. Because of its narrow therapeutic range, it is necessary to frequently monitor its concentration. We report the case of a 25-year-old man who underwent kidney transplantation whose tacrolimus concentrations, as measured by an affinity column-mediated immunoassay, were falsely elevated. As we reduced the dose of tacrolimus, the recipient developed T cell-mediated rejection. Using the same blood samples, an enzyme-multiplied immunoassay technique showed that the patient's levels of tacrolimus were extremely low. A further examination indicated that the false increase in the tacrolimus concentration was likely due to an unknown interfering substance. We administered methylprednisolone and antithymocyte-globulin. The patient's serum creatinine level decreased and remained stable after these treatments.

Content from these authors
© 2018 by The Japanese Society of Internal Medicine
Previous article Next article
feedback
Top