A Low Critical Event Rate Despite a High Abnormal Event Rate in Patients with Cardiac Implantable Electric Devices Followed Up by Remote Monitoring
Yoshimasa MorimotoNobuhiro NishiiSaori TsukudaSatoshi KawadaMasakazu MiyamotoAkihito MiyoshiKoji NakagawaAtsuyuki WatanabeKazufumi NakamuraHiroshi MoritaHiroshi Ito
Author information
JOURNAL OPEN ACCESS

2019 Volume 58 Issue 16 Pages 2333-2340

Details
Abstract

Objective Remote monitoring (RM) of cardiac implantable electric devices (CIEDs) has been advocated as a healthcare standard. However, expert consensus statements suggest that all patients require annual face-to-face follow-up consultations at outpatient clinics even if RM reveals no episodes. The objective of this study was to determine the critical event rate after CIED implantation through RM.

Methods This multicenter, retrospective, cohort study evaluated patients with pacemakers (PMs), implantable cardioverter defibrillators (ICDs), or cardiac resynchronization therapy defibrillator (CRT-Ds) and analyzed whether or not the data drawn from RM included abnormal or critical events.

Patients A total of 1,849 CIED patients in 12 hospitals who were followed up by the RM center in Okayama University Hospital were included in this study.

Results During the mean follow-up period of 774.9 days, 16,560 transmissions were analyzed, of which 11,040 (66.7%) were abnormal events and only 676 (4.1%) were critical events. The critical event rate in the PM group was significantly lower than that in the ICD or CRT-D groups (0.9% vs. 5.0% or 5.9%, p<0.001). A multivariate analysis revealed that ICD, CRT-D, and a low ejection fraction were independently associated with critical events. In patients with ICD, the independent risk factors for a critical event were old age, low ejection fraction, Brugada syndrome, dilated phase hypertrophic cardiomyopathy and arrhythmogenic right ventricular cardiomyopathy.

Conclusion Although abnormal events were observed in two-thirds of the transmitted RM data, the critical event rate was <1% in patients with a PM, which was lower in comparison to the rates in patients with ICDs or CRT-Ds. A low ejection fraction was an independent predictor of critical events.