日本鑑識科学技術学会誌
Online ISSN : 1882-2827
Print ISSN : 1342-8713
ISSN-L : 1342-8713
技術報告
ヘモキャッチを用いた人血検査に及ぼす界面活性剤・消毒薬の影響
白石 智子関口 和正大森 毅櫻田 宏一
著者情報
ジャーナル フリー

2004 年 9 巻 2 号 p. 143-149

詳細
抄録

  OC-Hemocatch, an immunochromatographic test device for fecal occult blood, has been used for the forensic identification of human blood. False-positive and false-negative effects by detergents and disinfectants on detection of human blood using OC-Hemocatch were investigated in this study. Dishwashing and laundry detergents showed no false-positive but false-negative results. Four in 8 disinfectants showed false-positive results, and 3 in the other 4 disinfectants inhibited the detection of human blood. Cationic and anionic surfactants showed false-positive results, but non-ionic and amphoteric surfactants did not, suggesting that the antibody-conjugated blue latex particles were aggregated electrically by cations or anions and caught nonspecifically onto the detection line. False-negative results might be caused by denaturation or degradation of hemoglobin by the comtaminants.

著者関連情報
© 2004 日本法科学技術学会