Experimental Study
Tissue Harmonic Imaging
- Experimental Analysis of the Mechanism of Image Improvement -
Kazuaki TanabeMarek BelohlavekJames F GreenleafJames B Seward
Author information
JOURNAL FREE ACCESS

2000 Volume 64 Issue 3 Pages 202-206

Details
Abstract

Tissue harmonic scanning visually improves echocardiographic image quality. The aim of the present study was to objectively assess the improvement in harmonic image quality under controlled laboratory conditions. A tissue-mimicking phantom that contained 8-mm-diameter cystic lesions at depths ranging from 2 to 12 cm was used. Harmonic scans (1.7 MHz transmit, 3.4 MHz receive) of the phantom were obtained and lesion detectability was compared to that in scans acquired with 2 fundamental frequencies (2.0 and 3.3 MHz). A 2 cm-thick ethanol layer was also used to simulate the nonlinear effect of human fat. Cyst detectability was quantified by measurement of the contrast-to-speckle ratio (CSR). The results indicated no significant difference in the CSR between harmonic and fundamental images obtained without the ethanol layer. With images obtained with the ethanol layer, a relative increase of the CSR during harmonic imaging was observed with respect to fundamental imaging (p<0.05). In conclusion, a fat layer, here simulated by ethanol, plays a significant role in determining the resulting image quality. Without this layer, the contribution of the second harmonic mode was not significant. Thus, in a slim patient, the harmonic mode may not be as beneficial to image improvement as in an obese patient.

Content from these authors
© 2000 THE JAPANESE CIRCULATION SOCIETY
Previous article Next article
feedback
Top