保健物理
Online ISSN : 1884-7560
Print ISSN : 0367-6110
ISSN-L : 0367-6110
Response of Thermoluminescence Dosimeter CaF2: Tm to Low-Level Proton Beams Using Rutherford Backscattering
Pin-Chieh HSULi-Chen TSAIHuan NIUPao-Shan WENG
著者情報
キーワード: proton beam, TLD, glow curve
ジャーナル フリー

2001 年 36 巻 2 号 p. 131-136

詳細
抄録

Low-level proton beams from a van de Graaff accelerator operated at 1-3MeV with proton dose from 1.48 to 94.56mGy using the Rutherford backscattering method were used to study the response of thermoluminescence dosimeter CaF2: Tm. The glow curves, TL output and sensitivity obtained were analyzed with respect to proton energy, dose and dose rate. The experimental results indicate that TLD CaF2: Tm may also be used as a proton dosimeter in the low energy range.

著者関連情報
© Japan Health Physics Society
前の記事 次の記事
feedback
Top