Journal of Health Science
Online ISSN : 1347-5207
Print ISSN : 1344-9702
ISSN-L : 1344-9702
Glucose Determination Using a Flow System with Enzyme Reactor and Application to Analysis of Glucose Content in Beverages
Hisakazu MoriMikiko IshidaSumie Okamoto
著者情報
キーワード: beverage
ジャーナル フリー

1999 年 45 巻 3 号 p. 126-129

詳細
抄録
Glucose was determined using an apparatus containing an enzyme reactor in a flow line. The enzymes used for glucose assay were hexokinase and glucose-6-phosphate dehydrogenase. NADH formed by enzymatic reactions was fluorometrically detected. The optimal concentrations of ATP and NAD<+> in the carrier were determined. Of the buffers examined for use as a carrier medium, triethanolamine buffer was found to be the most favorable and the optimal pH of this buffer was pH 7.0. Under these optimal conditions, the calibration curve for glucose was linear in the range of 2.0-200μM(r=0.999). Detection limit was 1.7μM(S/N=3).This method was applied to the analysis of glucose in several beverages. Glucose content determined by the present method agreed with that determined by a commercially available test-kit method. The methods for the preparation of immobilized enzymes were examined to retain the activities of enzymes.
著者関連情報
© The Pharmaceutical Society of Japan
前の記事 次の記事
feedback