医学物理
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
資料
放射線治療施設レベルと精度管理に必要な機器
隅田 伊織小泉 雅彦岡本 裕之阿部 容久伊丹 純小高 喜久雄川村 慎二宇野 隆齋藤 秀敏遠藤 真広
著者情報
ジャーナル フリー

2011 年 31 巻 3 号 p. 84-89

詳細
抄録

Radiotherapy institutions carry out various types of treatments such as conventional radiation treatment, stereotactic radiotherapy (SRT), intensity-modulated radiation therapy (IMRT) and image-guided radiotherapy (IGRT). According to the radiation treatment type, various kinds of quality assurance (QA) equipment are used to maintain the accurate dose delivery to a patient with accurate positioning. The aim of this report was to propose the QA equipment required at various institutional levels. The institution levels were categorized according to the radiation treatment cases. Institutions in the basic category used conventional radiotherapy techniques employing field sizes larger than 4 cm×4 cm, or they had no full-time radiation oncologist. In addition to the basic level, five other categories were defined which were based on being able to carry out total body irradiation, three-dimensional conformal radiation therapy / SRT, IMRT, and IGRT. Minimum QA equipment requirements were proposed for each level. This proposal will facilitate the purchase of QA equipment for the installation or refurbishing of a linear accelerator.

著者関連情報
© 2011 日本医学物理学会
前の記事
feedback