Feasibility study on a noninvasive measurement system for boron concentration
Masayori IshikawaMasahiko HirasawaTakehiro TomitaniHideo MurayamaMasaharu Hoshi
Author information
JOURNAL FREE ACCESS

2005 Volume 25 Issue 1 Pages 3-12

Details
Abstract
The kinematics of boron compounds in vivo is crucial to fully and safely utilize boron neutron capture therapy. In this paper, we proposed a prompt gamma-ray Compton scatter camera (PG-CSC), which can evaluate the 3-D kinematics of boron compounds without invasion. This PG-CSC combines a Compton camera with prompt gamma-ray analysis. The results of its design optimization and tests using simulations showed that one-hour-measurement was sufficient to evaluate a 3-D boron concentration distribution in a rat's brain with the resolution of 1mm in FWHM.