原著
プラスチックシンチレータを液体シンチレーションカウンターに用いた14C と 35S の識別
古田 悦子横田 繁昭渡辺 裕夫
著者情報
ジャーナル フリー

2009 年 8 巻 1 号 p. 51-58

詳細
抄録

  Spectra for pure-beta nuclides with liquid scintillator by using a liquid scintillation counter (LSC) are usually shifted to low energy region caused by optical and chemical quenching processes. So the end-point of LSC spectrum does not show the theoretical maximum beta-ray energy (Emax). On the other hand, the end points of the spectra of 14C and 35S showed their Emax, when a plastic scintillator (PS) was used for measurement as a substitute for a liquid scintillator. Therefore, the possibility of identification of 14C and 35S with plastic scintillator applied to conventional liquid scintillation counter is the aim of this study. Spectra could be measured with good reproducibility. Moreover the reusability of the PS rinsed within a day or covered with a thin film was also confirmed.

著者関連情報
© 2009 日本放射線安全管理学会